Human PM20D1 ELISA Kit
★ Download MSDS ★
Assay range: 31.2 - 2000 pg/mL
Kit Size: 96 wells/kit
Other Names: Peptidase M20 domain containing 1, Cps1
INTRODUCTION
PM20D1 is a bidirectional N-fatty-acyl amino acid synthase/hydrolase that regulates the production of N-fatty-acyl amino acids. These metabolites are endogenous chemical uncouplers of mitochondrial respiration. In a UCP1-independent manner, maybe through interaction with mitochondrial transporters, they promote proton leakage into the mitochondrial matrix. PM20D1 may indirectly regulate the bodily dissipation of chemical energy as heat through thermogenic respiration.
PRINCIPLE OF THE ASSAY
This assay is a quantitative sandwich ELISA. The microplate is pre-coated with a polyclonal antibody specific for human PM20D1. Standards and samples are pipetted into the wells and any human PM20D1 present is bound by the immobilized antibody. After washing away any unbound substances, a horseradish peroxidase biotin-labelled polyclonal antibody specific for human PM20D1 is added to the wells. After wash step to remove any unbound reagents, streptavidin-horseradish peroxidase (STP-HRP) conjugate is added. After the last wash step, an HRP-substrate (TMB) solution is added and colour develops in proportion to the amount of human PM20D1 bound initially. The assay is stopped and the optical density of the wells determined using a microplate reader. Since the increases in absorbance are directly proportional to the amount of captured human PM20D1, the unknown sample concentration can be interpolated from a reference curve included in each assay.
ASSAY PERFORMANCE
A. Typical representation of standard curve
The following standard curve is provided for demonstration only. A standard curve should be generated for each set of sample assay.
Human PM20D1 (pg/mL)
Absorbance
(450 nm)
Blanked Absorbance
0
0.103
0
31.2
0.126
0.023
62.5
0.161
0.058
125
0.237
0.134
250
0.421
0.318
500
0.747
0.644
1000
1.134
1.031
2000
2.12
2.017
B. Sensitivity
The lowest level of human PM20D1 that can be measured by this assay is 7.8 pg/mL.
C. Specificity
No cross reactivity or interference was observed when test with recombinant mouse PM20D1 protein.
D. Precision
Intra-assay Precision (Precision within an assay)
One sample of known concentration was tested 8 times on one plate.
Sample Mean (pg/mL) SD CV (%) 1 32723.26 1145.522 3.5 Inter-assay Precision (Precision between assays)
One sample of known concentration was tested in 8 separate assays.
Sample Mean (pg/mL) SD CV (%) 1 4568.75 160.22 3.5 E. Spike and Recovery
Serum samples were assayed by adding 90 µL of sample and 10 µL of spike stock solution calculated to yield the intended 0, 50, 400 or 800 pg/mL spike concentration.
Spike level
Expected
(pg/mL)
Observed
(pg/mL)
Recovery (%)
Low spike (50 pg/mL)
49.83
47.83
95
Medium spike
(400 pg/mL)396.5
357.33
90.1
High spike (800 pg/mL)
679.83
724
106.5
F. Linearity and Recovery
To assess the linearity of the assay, samples containing high concentrations of human PM20D1 were serially diluted with the 1×Assay buffer to produce samples with values within the dynamic range of the assay.
Dilution
Measured (pg/mL)
Expected (pg/mL)
Recovery (%)
1:100
3072.46
27850.9
110.3
1:200
30770
27850.9
110.5
PUBLICATIONS CITING THIS PRODUCT
- Yang R, Hu Y, Lee CH, Liu Y, Diaz-Canestro C, Fong CH, Lin H, Cheng KK, Pravelil AP, Song E, Lam KS. PM20D1 Is a Circulating Biomarker Closely Associated with Obesity, Insulin resistance and Metabolic Syndrome. European Journal of Endocrinology. 2021 Nov 1;1(aop).