β2-Microglobulin (B2M) Turbidimetric Immunoassay Kit

β2-Microglobulin (B2M) Turbidimetric Immunoassay Kit

SKU: 51680
  • INTRODUCTION

    β2-Microglobulin(β2-MG), is a small-molecule globulin produced by lymphocytes, platelets, and polymorphonuclear leukocytes, with a molecular weight of 11,800 and a single-chain polypeptide consisting of 99 amino acids. It is the β chain (light chain) portion of the human leukocyte antigen (HLA) on the cell surface. It contains a pair of disulfide bonds in the molecule and contains no sugar; similar to the structure of the immunoglobulin stable region. It is widely found in plasma, urine, cerebrospinal fluid, saliva, and colostrum. Normal human β2-microglobulin synthesis rate and release from the cell membrane are fairly constant. Β2-microglobulin can be freely filtered from the glomerulus, 99.9% is absorbed in the proximal tubules, and in renal tubular epithelial cells Decomposition and destruction; therefore, the emission of β2-microglobulin is very small under normal circumstances.

    β2-MG mainly reflects glomerular filtration dysfunction, urine β2-MG mainly reflects renal tubular reabsorption function impairment. It also can be used as a tumor marker for some people with blood cell cancers (multiple myeloma, ly -mphoma) to give information about their likely prognosis. The health refere- nce of β2-MG is 0.8-2 mg/L in blood and 0.11-0.32 mg/L in urine.

    IMD developed β2-MG PETIA kit uses its unique antibodies and antigens, with accuracy and specificity fully validated using international standard

  • PRINCIPLE OF THE ASSAY

    This assay is a turbidimetric immunoassay for the quantitative measurement of β2-MG in human serum and plasma. A standard or sample is added into a cuvette and mixed with the reaction buffer R1. After a short incubation, the test reagent R2, which is a suspension of microparticles coated with β2-MG antibodies, is added into the cuvette and mixed. The presence of β2-MG in the standard or sample causes the immune-particles to aggregate. The extent to which the microparticles aggregate is quantified by the amount of light scattering measured as absorbance by a chemistry analyzer. The concentration of β2-MG in unknown samples can be interpolated from a reference curve using the standards provided.

  • PERFORMANCE OF THE KIT

    TYPICAL STANDARD CURVE

    β2-MG  (mg/L)

    Absorbance (570 nm)

    0

    -0.0042

    0.2

    0.0055

    1

    0.0640

    2

    0.1217

    10

    0.4353

    20

    0.6265

  • ASSAY PROCEDURE

    Assay procedures may vary depending on the automated chemistry analyzer to be used. A general example of assay procedures is stated as follow:

    1. Dispense 180µl of R1 into a clean cuvette

    2. Add 2µl of sample and incubate at 37˚C for 5 minutes

    3. Further add 60µl of R2

    4. Read change of absorbance at Main Wavelength 570 nm for 8 minutes after the addition of R2

    5. Calculate the concentration of β2-Microglobulin in unknown sample by interpolation from a reference curve using the standards provided

  • PACKING SPECIFICATIONS

    Cat. No.

    Size

    Approximately tests

    51680-05

    R1: 15ml, R2: 5ml

    100

    51680-10

    R1: 30ml, R2: 10ml

    200

    51680-20

    R1: 60ml, R2: 20ml

    400

    51680-50

    R1: 150ml, R2: 50ml

    1000

    51680-100

    R1: 300ml, R2: 100ml

    2000

    For bulk order, please enquire

  • OTHER MATERIALS REQUIRED

    Product           

    Catalog           Number

    Format

    Method

    Calibrator                 

    51680-S1

    5 x 1ml

     

    Control

    51680-C1

    2 x 1ml

     

PriceFrom $220.00

The currency used on this website is United States Dollars (USD). All prices are displayed and transactions are processed in USD. For orders outside Hong Kong or if you would like to have other currencies available, please contact us for a quote.